Tipo de Pago: | L/C, T/T, D/P, Western Union |
---|---|
Cantidad de pedido mínima: | 1 |
Plazo de entrega: | 1 días |
Modelo: JYK-901A
Syringe Mode | Micro and continuous injection |
Syringes | 10mL, 20mL, 30mL, 50/60mL syringes, self-identified automatically. |
Syringe Brands | Up to 11 syringe brands could be preset. |
Optional Infusion mode | Drug weight mode, volume/time mode. |
FlowRate Range | 10mL syringe: 0.1 ~ 300 mL/h 20mL syringe: 0.1 ~ 400 mL/h 30mL syringe: 0.1 ~ 600 mL/h 50mL syringe: 0.1 ~ 1200 mL/h |
Increment | 0.1mL/h |
Flow Rate Accuracy | ≤ ± 2% (including accuracy ± 1% on mechanism) |
Volume Limit | 0.1~ 9999.9 mL and empty mode available |
Volume Infused Display | 0 ~ 9999.9 mL |
Purge Rate | 10mL syringe: 300 mL/h 20mL syringe: 400 mL/h 30mL syringe: 600 mL/h 50mL syringe: 1200 mL/h |
Bolus rate | 10mL syringe: 0.1 ~ 300 mL/h 20mL syringe: 0.1 ~ 400 mL/h 30mL syringe: 0.1 ~ 600 mL/h 50mL syringe: 0.1 ~ 1200 mL/h |
Optional Inject Rate Unit | mL/h, mg/h, µg/kg/min, mg/kg/min, µg/kg/h, mg/kg/h, and µg/h. |
KVO Rate | 1 mL/h |
Occlusion Pressure | Pressure detection device with 8 sensitivities setting, 20 ~ 140KPa, dynamically display pressure value. |
Audio and Visual Alarms | AC power lost, Barrel clasp, Plunger head, Finished. KVO, Near empty, Empty, Occlusion, Low battery, Bat. ran out, Battery lost, No operat. |
Display | Blue LCD with high brightness |
CommunicationPort | RS-232 (for nurse call) |
Power Source | AC: 100V ~ 240V, 50Hz/60Hz, ≤ 25VA |
Battery | Rechargeable Ni-MH battery: DC9.6V, 2000mAh, life: ≥ 4 hours at 5mL/h, remaining battery displayed. |
Operating Conditions | Temperature: 5°C ~ 40°C Relative humidity: 20% ~ 90% Atmospheric pressure: 70 ~ 106KPa |
Storage Conditions | Temperature: -20°C ~ 55°C Relative humidity: 10% ~ 90% Atmospheric pressure: 70 ~ 106KPa |
Classification | Class II, CF, IPX4 |
Dimensions and N.W. | 305mmx 185mm x 165mm; 2.0kg |
Application | Intravenous injection |
Grupos de Producto : Bomba de infusión y jeringa